  • Product Name: HX-902
  • Product Number: HX-902

 

The Features of Syringe Pump HX-902

1.         Complies with IEC60601,high safety designation with dual CPUs.

2.         Automatically identify size for syringes of 5mL, 10mL, 20mL, 30mL and 50/60mL.

3.         Available for Bolus mode.

4.         Available for drug weight mode.

5.         10 usual syringe brands preset, and easily customize for others.

6.         High brightness LCD display and user-friendly operation interface.

7.         Pressure detection device with 8 sensitivities setting, dynamically display pressure value.

8.         Pressure release function when occlusion occurs within the syringe.

9.         CE certified.

 

The Specifications of Syringe Pump HX-902

Syringe Mode

Micro and continuous injection

Syringes

5mL,10mL,20mL,30mL,50/60mL syringes,self-identified automatically

Syringe Brands

10 usual syringe brands preset, and easily customize for others

Flow Rate Range

5mL syringe: 0.1~60mL/h

10mL syringe:0.1~300mL/h    20mL syringe:0.1~400mL/h   

30mL syringe:0.1~600mL/h    50mL syringe:0.1~1200mL/h   

Increment

0.1 mL/h

Flow Rate Accuracy

≤±2% (including accuracy ±1% on mechanism)

Volume Limit

0.1~9999.9 mL and empty mode available

Volume Infused Display

0.1~9999.9 mL

Purge Rate

5mL syringe: 60mL/h

10mL syringe: 300mL/h    20mL syringe: 400mL/h   

30mL syringe: 600mL/h    50mL syringe: 1200mL/h   

Bolus Rate

5mL syringe: 60mL/h

10mL syringe:0.1~300mL/h    20mL syringe:0.1~400mL/h   

30mL syringe:0.1~600mL/h    50mL syringe:0.1~1200mL/h   

Optional Inject Rate Unit

mL/h, mg/h, µg/kg/min, mg/kg/min, µg/kg/h, mg/kg/h, and µg/h

KVO Rate

1 mL/h

Occlusion Pressure

Pressure detection device with 8 sensitivities setting,20~140Kpa,

Dynamically display pressure value

Audio and Visual Alarms

AC power disconnected, Syringe barrel clasp, Plunger head position

Injection finished, End of injection pre-alarm, Empty syringe, Occlusion, Low battery, Battery run out, Battery damaged, No operation, System failed

Display

Blue LCD with high brightness

Communication Port

RS-232(for nurse call)

Power Source

AC: 100V~240V, 50Hz/60Hz, 25VA

Battery

Rechargeable Ni-MH battery:DC9.6V,2000mAh,life:4 hours at 5mL/h, remaining battery displayed

Rechargeable Lithium battery:DC11.1V,2000mAh,life:4 hours at 5mL/h, remaining battery displayed

Operating Conditions

Temperature:5~40,Relative humidity:20%~90%,Atmospheric pressure:70~106KPa

Storage Conditions

Temperature:-20~40,Relative humidity:20%~90%,Atmospheric pressure:70~106KPa

Classification

Class II, CF, IPX4

Dimensions and Net Weight

335mm×220mm×205mm;2.0kg

Application

Intravenous injection